config:
theme: 'neutral'
---
flowchart LR
subgraph MRIRoom
MriMon[Monitor]
MriRespGrip[Response Grips]
MriScan[MRI Scanner]
subgraph fMRITrigger
fMriTrigI1[opt_trig_in]
fMriTrigO2[opt_trig_out_aux]
end
end
subgraph ControlRoom
subgraph StimulusComp
StimUsb[USB]
StimHdmi[HDMI]
end
subgraph SyncBox
subgraph UserInterface
SyncUsb[USB]
SyncTrig[fMRI Trigger]
SyncTtlIn[TTL]
end
subgraph ResponseCapture
SyncRespGrip[Response Grips]
SyncTtlOut[TTL]
end
end
subgraph VideoConverter
VidHdmi[HDMI In]
VidOptic[Optic Out]
end
end
MriScan -- W13470 --> fMriTrigI1
fMriTrigO2 --> SyncTrig
MriRespGrip --> SyncRespGrip
SyncTtlOut --> SyncTtlIn
SyncUsb <--> StimUsb
StimHdmi --> VidHdmi
VidOptic --> MriMon
style MRIRoom fill:#ffffff,stroke:#800080
style ControlRoom fill:#ffffff,stroke:#00FF00
style fMRITrigger fill:#ffffff,stroke:#008080
style StimulusComp fill:#ffffff,stroke:#0000FF
style SyncBox fill:#ffffff,stroke:#FF0000
style UserInterface fill:#ffffff,stroke:#000000
style ResponseCapture fill:#ffffff,stroke:#000000
style VideoConverter fill:#ffffff,stroke:#000000
Operate NNL SyncBox
Overview
The Nordic Neuro Lab (NNL) SyncBox helps align participant responses (real time) with fMRI acquistion (volume time). To do this, the SyncBox receives (a) synchronization pulse from the scanner and (b) participant responses. These are converted from fiber optic to serial output and sent to a computer via USB.
Unlike more common synchronization boxes, the NNL SyncBox does not passively listen for input once a session is begun. Rather, the user most program the number of volumes and TR length for each session. This can be done in Manual or Controlled Mode.
SyncBox Operations
- The NNL SyncBox operates as a serial device. Install this driver on the device connecting to the SyncBox (e.g. stimulus computer).
- Use the USB cable on the left side of SyncBox to connect with the stimulus computer.
- Power the SyncBox on with the toggles on the right side.
Manual Mode
Two manual modes exist, Simulation and Synchronization. These are identical, save (a) Simulation generates faux synchronization pulses, and (b) Synchronization listens for the scanner synchronization pulse.
- Setup:
- Verify Serial mode: Sync/Simulation -> Go to Options menu -> USB Mode -> Serial.
- HID is still an option (but does not work).
- Manually send a sync pulse to your program (for testing): Sync/Simulation -> Send triggerpulse to PC.
- Verify Serial mode: Sync/Simulation -> Go to Options menu -> USB Mode -> Serial.
- Use:
- Configure session by specifying TR length and number of volumes.
- Note: the SyncBox will stop listening after the specified number of volumes.
- Start the box listening for synchronization pulse and/or response grip input: Sync/Simuation -> Start Session.
- Use center button to exit Session.
- Configure session by specifying TR length and number of volumes.
Controlled Mode
- This is the vendor preferred method of using the SyncBox.
- An example of using Python to connect the SyncBox to PsychoPy is available here.
SyncBox Configuration
Front
- The SyncBox consists of the User Interface (A) and Response Capture (B).
- The User Interface provides a navigation pad, LCD screen, and USB output for a task computer.
- The Response Capture contains one green power light and four response lights.
The Response Capture mappings are:
Light Grip Button Character 1 L. Thumb A 2 L. Index B 3 R. Index C 4 R. Thumb D
Back
- The User Interface receives input from the fMRI Trigger Converter box (A1), Response Capture (A2), and Power (A3).
- The Response Capture receives input from the Response Grips (B1) and Power (B3), and sends output to the User Interface (B2).
Video Converter Configuration
Video from the stimulus computer is sent to the presentation monitor in Zone 4 via the 4K HDMI to Fiber Extender. While this box is capable of bidirectional communication (Tx/Rx), the only ‘downstream’ device is the monitor.
- A. USB, unused but could send/receive serial data (keyboard, mouse).
- B. Fiber output, connects to the presentation monitor.
- C. HDMI input, from the stimulus computer.
- D. Power.
fMRI Trigger Configuration
- The Siemens fMRI Trigger Converter box is located in the MRI Equipment Room (B18A).
- The fMRI trigger box receives a synchronization pulse as fiber optic (Cable W13470) input to the port “opt_trig_in” (I1). It is powered by a 5V 500mA adapter via a USB B-Type cable (I2).
- The fMRI trigger box then sends output from ports “opt_trig_out” (Cable W13471; O1) and “opt_trig_out_aux” (O2). When the fMRI trigger box is in “impulse” mode a synchronization pulse is sent for every volume, and every other volume in “toggle” mode.
- The fiber optic cable connected to “opt_trig_out_aux” (O2) is the input for the NNL SyncBox (A1).